Two calibration points will be fine, pH 4 and pH 7. Get some glass wear from the kitchen and add some reference solution. Put the pH or ORP probe in the solution and wait for the value to adjust, it usually takes one or two minutes. You can use the serial monitor on the Nanode ( un-comment out the #debug serial.print function) to read the output of your pH and ORP probe. After one point calibration, spray or rinse the glass junction with distilled water for a couple minutes before putting it in the other standard solution.
The ORP reading for the standard reference solution is as follows:
pH Value | mV Value | pH Value | mV Value | pH Value | mV Value | pH Value | mV Value | pH Value | mV Value |
---|---|---|---|---|---|---|---|---|---|
0 | 990 | 3 | 800 | 6 | 640 | 9 | 460 | 12 | 280 |
1 | 920 | 4 | 740 | 7 | 580 | 10 | 400 | 13 | 220 |
2 | 860 | 5 | 680 | 8 | 520 | 11 | 340 | 14 | 160 |
The ORP probe in the pH 4 reference solutions should read around 740 mV and the pH 7 reference solution should read around 580 mV.
No comments:
Post a Comment